User Tools

Site Tools


oxford_hypersense

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
oxford_hypersense [2010/02/02 16:15]
miquel
oxford_hypersense [2012/02/17 13:11] (current)
miquel [Agraïments]
Line 1: Line 1:
 ====== Hiperpolaritzador ====== ====== Hiperpolaritzador ======
  
-{{:administracio_sermn:oxford_dnp:oxfory-hypersense.png?200 | Oxford Hypersense}} L'hiperpolaritzador **Oxford Hypersense** es va adquirir a mitjans de 2009, va arribar a la UAB a principis de febrer de 2010, i la seva instal·lació es va programar per la primavera de 2010. Amb l'hiperpolaritzador va adquirir-se també un polarímetre 13C-MQC Polarimeter amb què mesurar el nivell d'hiperpolarització de la mostra.+{{:imatges:equip_cientific:oxfory-hypersense.png?200 | Oxford Hypersense}} L'hiperpolaritzador **Oxford Hypersense** es va adquirir a mitjans de 2009, va arribar a la UAB a principis de febrer de 2010 ([[https://sermn02.uab.cat/sermn-pictures/main.php?g2_itemId=1060 | veure fotos]]), i la seva instal·lació es va programar per la primavera de 2010. Amb l'hiperpolaritzador va adquirir-se també un polarímetre 13C-MQC Polarimeter amb què mesurar el nivell d'hiperpolarització de la mostra.
  
-Per una descripció breu de l'aparell consulta els documents {{:administracio_sermn:oxford_dnp:oxford_instruments_-_hypersense_brochure_-_03-2007.pdf|Hypersense Brochure}} i {{:administracio_sermn:oxford_dnp:oxford_-_13c-mqc_polarimeter.pdf|13C-MQC Polarimeter}}+Per una descripció breu de l'aparell consulta els documents {{:oxford_dnp:oxford_instruments_-_hypersense_brochure_-_03-2007.pdf | Hypersense Brochure}} i {{:oxford_dnp:oxford_-_13c-mqc_polarimeter.pdf | 13C-MQC Polarimeter}}
  
 Per conèixer les aplicacions que poden beneficiar-se de l'ús d'un hiperpolaritzador consulta la informació (pòsters, presentacions, articles i demés) disponibles a l'[[http://www.oxinst.com/products/dnp/applications-resource/Pages/applications-library.aspx | Application Library]] d'Oxford Instruments. Per conèixer les aplicacions que poden beneficiar-se de l'ús d'un hiperpolaritzador consulta la informació (pòsters, presentacions, articles i demés) disponibles a l'[[http://www.oxinst.com/products/dnp/applications-resource/Pages/applications-library.aspx | Application Library]] d'Oxford Instruments.
  
 +^   Technical Specifications   ^^
 +^ Microwave source          | Integrated 94 GHz with 0.5 GHz sweep (user selectable) up to 100 mW   |
 +^ Magnetic field strength   | 3.35 T actively shielded    |
 +^ Sample temperature        | Selectable <3.9 K,  ≥1.4 K **(1)** |
 +^ Helium refill volume      | 65 L from minimum level   |
 +^ Nitrogen refill volume    | 60 L from minimum level   |
 +^ Helium consumption during dissolutions    | Typical <2 L per dissolution **(2)**   |
 +^ Minimun operating ceiling height          | <3.0 m   |
 +^ System weight including cryogens          | approx. 600 kg  |
 +^ Trigger output for NMR spectrometer       | 5 V signal    |
 +^ Heat dissipation          | 3 kW  (vacuum pump included   |
 +^ Dimensions - Footprint    | W 1668 mm (~ 66") x D 1038 mm (~ 41"  |
 +^ Dimensions - Height       | 2860 mm (~ 113"  |
 +| **(1)** As measured by instrument control system with 65% helium insert level at microwave power 100 mW over 30 minutes.   ||
 +| **(2)** Averaged over 10 repeatable runs with a temperature of 1.4 K over a 6 hour polarisation and dissolution, 100 mW microwave power with helium level at 65%.   ||
  
  
Line 13: Line 28:
 La informació sobre la compra i configuració de l'hiperpolaritzador, així com les incidències relacionades amb aquest instrument només són accessibles al personal del SeRMN i es troben a la [[:administracio_sermn:oxford_hypersense | intrawiki]] (cal login). La informació sobre la compra i configuració de l'hiperpolaritzador, així com les incidències relacionades amb aquest instrument només són accessibles al personal del SeRMN i es troben a la [[:administracio_sermn:oxford_hypersense | intrawiki]] (cal login).
  
 +===== Agraïments =====
  
 +La compra d'aquest equip ha estat cofinançada pel [[http://ec.europa.eu/regional_policy/funds/feder/index_en.htm | Fons Europeu de Desenvolupament Regional (FEDER - Una forma de fer Europa).]]\\ The purchase of this equipment has been funded by the [[http://ec.europa.eu/regional_policy/funds/feder/index_en.htm | European Regional Development Fund (ERDF - A way of making Europe).]]
 +
 +{{:logos:feder_funded_equipment.png?200|FEDER Funded Equipment}}
  
oxford_hypersense.1265123720.txt.gz · Last modified: 2010/02/02 16:15 by miquel