 +==== Fitxer "uxnmr.info" ====
 +
 +<file>
 +# cat uxnmr.info 
 +CONFIGURATION INFORMATION
 +=========================
 +
 +Path         : /opt/topspin3.6.4/conf/instr/spect/uxnmr.info
 +Date         : Thu Aug 26 16:17:50 2021
 +Release      : TopSpin 3.6.4
 +Installed in : /opt/topspin3.6.4
 +Host         : cie-54-214.uab.cat
 +OS           : CentOS Linux release 7.9.2009 (Core)
 +CPU          : Intel(R) Xeon(R) CPU E3-1246 v3 @ 3.50GHz (8 cores at 3500 MHz with Hyperthreading)
 +User         : sermnadmin (SeRMN Staff)
 +System       : Avance III 400 NMR spectrometer
 +1H-frequency : 400.13 MHz
 +Description  : Avance III 400 MHz SB - SeRMN, UAB
 +Bruker Order : Instal·lat al 1993
 +Configured in: cie-54-214.uab.cat:/opt/topspin3.6.4/conf/instr/spect
 +
 +IPSO: connected to spectrometer subnet
 +- TCP/IP address = 149.236.99.90
 + - IPSO type: AQS
 +- Tctrl : 1
 +- Fctrls: 2
 +- Gctrl1: with digital preemphasis
 +- Rctrl : 1
 +- FREDs : none
 +- DPP : 1
 +
 +DRU1: AQS DRU Z100977/00582 ECL 04.00
 +- TCP/IP address = 149.236.99.89
 +- Firmware Version = 170619
 +- DRU1 controls AQS-Rack and HPPR/2
 +
 +AQS: connected to 149.236.99.89:/dev/tty10
 +  _Slot_ SBSB _____________________Board_____________________
 +  Number Addr Type HW-VS FW-VS ID  ECL Name   Description
 +  -----------------------------------------------------------
 +       0x10 0x72   0x3    AW  R  3.0 REC-1  AQS RXAD400 Z102116/705 ECL 03.00
 +       0x34  0x2   0x1        X  5.0 REF-1  REF-400 Reference Board for AQS Receiver
 +       0x24 0x11   0x8    AE  S  2.0 SGU-1  AQS SGU/2 400 Z103080/00351 ECL 02.00 FPGA-Vs=AF
 +       0x25 0x11   0x8    AE  S  2.0 SGU-2  AQS SGU/2 400 Z103080/00340 ECL 02.00 FPGA-Vs=AF
 +       0x36  0x5            A  6.0 BLA-1  BLA2BB 150/60 20-400 W1345049/1817 ECL 6
 +    21   0x21 0xcf            P  1.0 PSD-1  PSD/3 Power Supply Distribution Board
 +    --   0x20 0xd1            B  2.0 MASTER AQS/3 Chassis
 +       0x20  0x7            B      MASTER AQS Rack Master
 +  Logical Receiver/SGU configuration:
 +
 +Router: 1 AQS-Minirouter
 +
 +Amplifiers (AQS configuration):
 +  RO  Amplifier  Module  Nucleus  Power/ Switchbox Name PN/SN/FW
 +        1                B      150    N         BLA2BB 150/60 20-400 W1345049/1817/-
 +        1                B       60    N         BLA2BB 150/60 20-400 W1345049/1817/-
 +        2               2H       20    N         2HTX in BSMS -/-/-
 +
 +Gradient amplifiers at the spectrometer subnet:
 +----------------------------------------------
 +BGA1: BGA_W1213891_0040
 +- TCP/IP address   = 149.236.99.92
 +- Firmware version = 20080206
 +- Web version      = 1.1
 +- Current limits   = 0.0/X, 0.0/Y, -10.0/Z (in A)
 +
 +BSMS: BSMS/2 connected to ethernet
 +- TCP/IP address = 149.236.99.20
 +- ELCB firmware version = 20150423
 +- SCB channels = 20
 +- Shim System = BOSS1-SB
 +- 2HTX Amplifier: available
 +- Lock: installed
 +
 +MAS2 control unit: connected to 149.236.99.90:/dev/tty05
 +- Boot Software        : 20031216
 +- Application Software : 20120329
 +- Insert Time          : 10 s
 +- Eject Time           : 10 s
 +- Spin Lock Tolerance  : 5 Hz
 +
 +Eurotherm VTU: device connected to 149.236.99.90:/dev/tty04
 +
 +Preamplifiers :
 +HPPR2: - HPPR/2 preamplifier connected to 149.236.99.89:/dev/tty10
 +    Type      : HPPR/2
 +    Controller: Cover/1 with Application firmware = AQ
 +                15 LEDs for tuning, 15 LEDs for matching
 +    Module  1 : 1H/2H 1H (virtual 50 Ohm reference: 274.9%/85.9deg)
 +                PN=Z003959, SN=00418 from 20070319
 +    Module  2 : 1H/2H 2H (virtual 50 Ohm reference: 116.6%/-71.1deg)
 +                PN=Z003959, SN=00418 from 20070319
 +    Module  3 : XBB19F 2HS
 +                PN=Z003504, SN=00415 from 20070425
 +
 +Gradient Controller cable connections
 +--------------------------------------------------------------------
 +GCtrl1: connected via DPP to GREAT 1-10 Gradient Amplifier Unit W1213891/0040
 +
 +RF cable connections (assumed)
 +--------------------------------------------------------------------
 +SGU1 NORM output -> input 2 of transmitter 1 (BLA2BB 150/60 20-400 W1345049/1817)
 +SGU1 AUX  output -> input 1 of transmitter 2 (2HTX in BSMS)
 +SGU2 NORM output -> input 3 of transmitter 1 (BLA2BB 150/60 20-400 W1345049/1817)
 +SGU2 AUX  output -> TUNE signal input of HPPR
 +
 +Blanking cable connections (assumed)
 +--------------------------------------------------------------------
 +transmitter 1 = BLA2BB 150/60 20-400 W1345049/1817:
 +- amplifier B-150W uses blanking 1
 +- amplifier B-60W uses blanking 2
 +
 +transmitter 2 = 2HTX in BSMS:
 +- amplifier 2H-20W uses blanking 3
 +</file>
 +
 +==== Fitxers de configuració de les connexions RS232 ====
 +
 +La configuració de les connexions per port sèrie (RS232) és la mateixa que a la versió 3.6.3 perquè he copiat els fitxers de configuració,
 +
 +  * ''~topspin/conf/instr/spect/rs232_device/aqs''
 +  * ''~topspin/conf/instr/spect/rs232_device/mas''
 +  * ''~topspin/conf/instr/spect/rs232_device/mas2''
 +  * ''~topspin/conf/instr/spect/rs232_device/preamp1''
 +  * ''~topspin/conf/instr/spect/rs232_device/temp''
  
  
informatica/hp_z230_400sb_topspin364_1.1629987724.txt.gz · Last modified: 2021/08/26 16:22 by miquel