User Tools

Site Tools


informatica:hp_z230_400sb_topspin364_1

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
Next revision
Previous revision
Last revision Both sides next revision
informatica:hp_z230_400sb_topspin364_1 [2021/08/26 16:22]
miquel [Configuració de l'espectròmetre]
informatica:hp_z230_400sb_topspin364_1 [2021/08/26 16:31]
miquel [Fitxers de configuració de les connexions RS232]
Line 276: Line 276:
 </WRAP> </WRAP>
  
 +==== Fitxer "uxnmr.info" ====
 +
 +<file>
 +# cat uxnmr.info 
 +CONFIGURATION INFORMATION
 +=========================
 +
 +Path         : /opt/topspin3.6.4/conf/instr/spect/uxnmr.info
 +Date         : Thu Aug 26 16:17:50 2021
 +Release      : TopSpin 3.6.4
 +Installed in : /opt/topspin3.6.4
 +Host         : cie-54-214.uab.cat
 +OS           : CentOS Linux release 7.9.2009 (Core)
 +CPU          : Intel(R) Xeon(R) CPU E3-1246 v3 @ 3.50GHz (8 cores at 3500 MHz with Hyperthreading)
 +User         : sermnadmin (SeRMN Staff)
 +System       : Avance III 400 NMR spectrometer
 +1H-frequency : 400.13 MHz
 +Description  : Avance III 400 MHz SB - SeRMN, UAB
 +Bruker Order : Instal·lat al 1993
 +Configured in: cie-54-214.uab.cat:/opt/topspin3.6.4/conf/instr/spect
 +
 +IPSO: connected to spectrometer subnet
 +- TCP/IP address = 149.236.99.90
 + - IPSO type: AQS
 +- Tctrl : 1
 +- Fctrls: 2
 +- Gctrl1: with digital preemphasis
 +- Rctrl : 1
 +- FREDs : none
 +- DPP : 1
 +
 +DRU1: AQS DRU Z100977/00582 ECL 04.00
 +- TCP/IP address = 149.236.99.89
 +- Firmware Version = 170619
 +- DRU1 controls AQS-Rack and HPPR/2
 +
 +AQS: connected to 149.236.99.89:/dev/tty10
 +  _Slot_ SBSB _____________________Board_____________________
 +  Number Addr Type HW-VS FW-VS ID  ECL Name   Description
 +  -----------------------------------------------------------
 +       0x10 0x72   0x3    AW  R  3.0 REC-1  AQS RXAD400 Z102116/705 ECL 03.00
 +       0x34  0x2   0x1        X  5.0 REF-1  REF-400 Reference Board for AQS Receiver
 +       0x24 0x11   0x8    AE  S  2.0 SGU-1  AQS SGU/2 400 Z103080/00351 ECL 02.00 FPGA-Vs=AF
 +       0x25 0x11   0x8    AE  S  2.0 SGU-2  AQS SGU/2 400 Z103080/00340 ECL 02.00 FPGA-Vs=AF
 +       0x36  0x5            A  6.0 BLA-1  BLA2BB 150/60 20-400 W1345049/1817 ECL 6
 +    21   0x21 0xcf            P  1.0 PSD-1  PSD/3 Power Supply Distribution Board
 +    --   0x20 0xd1            B  2.0 MASTER AQS/3 Chassis
 +       0x20  0x7            B      MASTER AQS Rack Master
 +  Logical Receiver/SGU configuration:
 +
 +Router: 1 AQS-Minirouter
 +
 +Amplifiers (AQS configuration):
 +  RO  Amplifier  Module  Nucleus  Power/ Switchbox Name PN/SN/FW
 +        1                B      150    N         BLA2BB 150/60 20-400 W1345049/1817/-
 +        1                B       60    N         BLA2BB 150/60 20-400 W1345049/1817/-
 +        2               2H       20    N         2HTX in BSMS -/-/-
 +
 +Gradient amplifiers at the spectrometer subnet:
 +----------------------------------------------
 +BGA1: BGA_W1213891_0040
 +- TCP/IP address   = 149.236.99.92
 +- Firmware version = 20080206
 +- Web version      = 1.1
 +- Current limits   = 0.0/X, 0.0/Y, -10.0/Z (in A)
 +
 +BSMS: BSMS/2 connected to ethernet
 +- TCP/IP address = 149.236.99.20
 +- ELCB firmware version = 20150423
 +- SCB channels = 20
 +- Shim System = BOSS1-SB
 +- 2HTX Amplifier: available
 +- Lock: installed
 +
 +MAS2 control unit: connected to 149.236.99.90:/dev/tty05
 +- Boot Software        : 20031216
 +- Application Software : 20120329
 +- Insert Time          : 10 s
 +- Eject Time           : 10 s
 +- Spin Lock Tolerance  : 5 Hz
 +
 +Eurotherm VTU: device connected to 149.236.99.90:/dev/tty04
 +
 +Preamplifiers :
 +HPPR2: - HPPR/2 preamplifier connected to 149.236.99.89:/dev/tty10
 +    Type      : HPPR/2
 +    Controller: Cover/1 with Application firmware = AQ
 +                15 LEDs for tuning, 15 LEDs for matching
 +    Module  1 : 1H/2H 1H (virtual 50 Ohm reference: 274.9%/85.9deg)
 +                PN=Z003959, SN=00418 from 20070319
 +    Module  2 : 1H/2H 2H (virtual 50 Ohm reference: 116.6%/-71.1deg)
 +                PN=Z003959, SN=00418 from 20070319
 +    Module  3 : XBB19F 2HS
 +                PN=Z003504, SN=00415 from 20070425
 +
 +Gradient Controller cable connections
 +--------------------------------------------------------------------
 +GCtrl1: connected via DPP to GREAT 1-10 Gradient Amplifier Unit W1213891/0040
 +
 +RF cable connections (assumed)
 +--------------------------------------------------------------------
 +SGU1 NORM output -> input 2 of transmitter 1 (BLA2BB 150/60 20-400 W1345049/1817)
 +SGU1 AUX  output -> input 1 of transmitter 2 (2HTX in BSMS)
 +SGU2 NORM output -> input 3 of transmitter 1 (BLA2BB 150/60 20-400 W1345049/1817)
 +SGU2 AUX  output -> TUNE signal input of HPPR
 +
 +Blanking cable connections (assumed)
 +--------------------------------------------------------------------
 +transmitter 1 = BLA2BB 150/60 20-400 W1345049/1817:
 +- amplifier B-150W uses blanking 1
 +- amplifier B-60W uses blanking 2
 +
 +transmitter 2 = 2HTX in BSMS:
 +- amplifier 2H-20W uses blanking 3
 +</file>
 +
 +==== Fitxers de configuració de les connexions RS232 ====
 +
 +La configuració de les connexions per port sèrie (RS232) és la mateixa que a la versió 3.6.3 perquè he copiat els fitxers de configuració,
 +
 +  * ''~topspin/conf/instr/spect/rs232_device/aqs''
 +  * ''~topspin/conf/instr/spect/rs232_device/mas''
 +  * ''~topspin/conf/instr/spect/rs232_device/mas2''
 +  * ''~topspin/conf/instr/spect/rs232_device/preamp1''
 +  * ''~topspin/conf/instr/spect/rs232_device/temp''
 +</file>
  
informatica/hp_z230_400sb_topspin364_1.txt · Last modified: 2021/08/26 16:31 by miquel